Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations
1-30 PerSeptive Biosystems
1
EMISTurns BA1 and BA2 on or off. Used during
troubleshooting only.
BA2 Pressure in mirror
chamber
(Voyager-DE PRO only)
Less than
2x10
-7
TC2 Pressure in sample loading
chamber
Less than 10
-2
during
operation.
Higher when
loading or
ejecting sample
plate.
TC1,
TC3,
TC4
Not used, displays E03
(indicates gauge not
connected)
______
Gauge Measures
Expected
Pressure
Vedere la pagina 59