Chapter 5 Optimizing Instrument Settings
5-70 PerSeptive Biosystems
5
NOTE: Changing the Delay Time in increments smaller
than 25 nsec may have no significant impact on
resolution. For information, see “Inherent Delay Time
offset” on page 5-67.
2,000–10,000 100–500 72–78
10,000–100,000 300–600 72–78
>100,000 No data
available
No data
available
Table 5-13 Voyager-DE PRO Reflector Mode—Delay
Time and Grid Voltage% Values (Continued)
Mass Range (Da)
Reflector Mode
Delay Time
(nsec)
Grid
Voltage%
Table 5-14 Voyager-DE STR Reflector Mode—Delay
Time and Grid Voltage% Values
Mass Range (Da)
Reflector Mode
Delay Time
(nsec)
Grid
Voltage%
500–2,000 50–100 70–80
2,000–10,000 50–500 70–80
10,000–20,000 200–700 70–80
20,000–100,000 500–1,000 70–80
>100,000 No data
available
No data
available